Now position:HomePRODUCTSSensing & ControlFlowmeters
NT® Electronic Flowmeter Model 5400

NT® Electronic Flowmeter Model 5400

● Provides simultaneous measurement of outlet pressure and flow rate.
● Ideal for capturing critical diagnostic information for monitoring or controlling process applications such as high-purity solvent dispense, precision blending and metering and totalized flow for custody transfer.
● Incorporates two pressure sensors, separated by a venturi style orifice, which provides simultaneous flow and pressure.
● No moving parts reduce particle generation potential.
● 316 stainless steel, PFA, and Kalrez® wetted surfaces for high-purity solvent applications.
● Integral pressure transducer provides additional process information.
● The flow-through design minimizes dead volume, reducing contamination potential.
● Available in two types of industry standard port connections: tube stub (swage style) type and face seal (metal gasket) type.
● 100% functionally tested prior to shipment. Performance is calibrated for an accuracy of 1% of full scale.


Download : 

 
 
Specifications :

Materials of construction:
  Wetted parts: Body: 316 stainless steel
Sensor interface (standard): PFA
Primary seal: Kalrez®
  Nonwetted parts: Anodized aluminum, FEP and Viton® (in addition to materials listed above)
Process temperature: 50° to 149°F (10° to 65°C) Consult factory for higher temperatures
Electrical input: 24 VDC (12-28 VDC input voltage)
Electrical output: Two 4-20 mA electrically isolated outputs, one for flow and one for outlet pressure
Pressure drop: 3 PSID at nominal flow, 10:1 turndown
12 PSID at nominal flow, 20:1 turndown
(Nominal flow = 80% of full scale flow)
Flow measurement accuracy:
  10:1 turndown: 20-100% of range ±1.0% FS
10-20% of range ±2.5% FS
  20:1 turndown: 10-100% of range ±1.0% FS
5-10% of range ±2.5% FS
(Accuracy stated as % of full scale, FS, using deionized water at 70°F [23°C] and includes the combined effects of linearity, hysteresis and repeatability)
Repeatability:
  10:1 turndown: 20-100% of range ±0.5% FS
10-20% of range ±1.0% FS
  20:1 turndown: 10-100% of range ±0.5% FS
5-10% of range ±1.0% FS
Operating pressure: 0 to 60 PSIG (0 to 414 kPa)
Over-pressure limit: 100 PSIG (690 kPa)
Pressure measurement accuracy: ±1% of full scale
(Includes combined effects of linearity, hysteresis and repeatability)
Electrical enclosure: NEMA 5/IP54